Patented Design
Gas analyzer and sonic anemometer in one sensor
Documentos relacionados WEATHER Documentos relacionados WATER Documentos relacionados ENERGY Documentos relacionados GAS FLUX & TURBULENCE Documentos relacionados INFRASTRUCTURE Documentos relacionados SOIL

Resumen

Campbell Scientific’s IRGASON fully integrates the open-path analyzer and sonic anemometer. Designed specifically for eddy-covariance flux measurements, the patented design is easier to install and use than separate sensors and provides increased measurement accuracy. The IRGASON simultaneously measures absolute carbon dioxide and water vapor, air temperature, barometric pressure, three-dimensional wind speed, and sonic air temperature. U.S. patent D680455

For more information about the benefits of having a co-located measurement, refer to the poster "Improved eddy flux measurements by open-path gas analyzer and sonic anemometer co-location."

Read More

Beneficios y características

  • Colocated measurement volumes for both analyzer and anemometer
  • Field serviceable (easy access to chemical bottles)
  • Unique optical configuration gives a slim aerodynamic shape with minimal wind distortion
  • Fully compatible with Campbell Scientific dataloggers; field setup, configuration, and field zero and span can be accomplished directly from the datalogger
  • Low power consumption; suitable for solar power applications
  • Low noise

Images

Descripción técnica

The IRGASON has the following outputs:

  • Ux (m/s)
  • Uy (m/s)
  • Uz (m/s)
  • Sonic Temperature (°C)
  • Sonic Diagnostic
  • CO2 Density (mg/m3)
  • H2O Density (g/m3)
  • Gas Analyzer Diagnostic
  • Ambient Temperature (°C)
  • Atmospheric Pressure (kPa)
  • CO2 Signal Strength
  • H2O Signal Strength
  • Source Temperature (°C)

Especificaciones

Patent U.S. Patent No. D680455
Operating Temperature Range -30° to +50°C
Calibrated Pressure Range 70 to 106 kPa
Input Voltage Range 10 to 16 Vdc
Power 5 W (steady state and power up) at 25°C
Measurement Rate 60 Hz
Output Bandwidth 5, 10, 12.5, or 20 Hz (user-programmable)
Output Options SDM, RS-485, USB, analog (CO2 and H2O only)
Auxiliary Inputs Air temperature and pressure
Warranty 3 years or 17,500 hours of operation (whichever comes first)
Cable Length 3 m (10 ft) from IRGASON to EC100
Weight
  • 3.2 kg (7.1 lb) for EC100 electronics
  • 2.8 kg (6.1 lb) for IRGASON head and cables

Gas Analyzer

Path Length 15.37 cm (6.05 in.)
A temperature of 20°C and pressure of 101.325 kPa was used to convert mass density to concentration.

Gas Analyzer - CO2 Performance

-NOTE- A temperature of 20°C and pressure of 101.325 kPa was used to convert mass density to concentration.
Accuracy
  • Assumes the following: the gas analyzer was properly zero and spanned using the appropriate standards; CO2 span concentration was 400 ppm; H2O span dewpoint was at 12°C (16.7 ppt); zero/span temperature was 25°C; zero/span pressure was 84 kPa; subsequent measurements made at or near the span concentration; temperature is not more than ±6°C from the zero/span temperature; and ambient temperature is within the gas analyzer operating temperature range.
  • 1% (standard deviation of calibration residuals)
Precision RMS (maximum) 0.2 mg/m3 (0.15 μmol/mol)

Nominal conditions for precision verification test: 25°C, 86 kPa, 400 μmol/mol CO2, 12°C dewpoint, and 20 Hz bandwidth.
Calibrated Range 0 to 1,000 μmol/mol (0 to 3,000 μmol/mol available upon request.)
Zero Drift with Temperature (maximum) ±0.55 mg/m3/°C (±0.3 μmol/mol/°C)
Gain Drift with Temperature (maximum) ±0.1% of reading/°C
Cross Sensitivity (maximum) ±1.1 x 10-4 mol CO2/mol H2O

Gas Analyzer - H2O Performance

-NOTE- A temperature of 20°C and pressure of 101.325 kPa was used to convert mass density to concentration.
Accuracy
  • Assumes the following: the gas analyzer was properly zero and spanned using the appropriate standards; CO2 span concentration was 400 ppm; H2O span dewpoint was at 12°C (16.7 ppt); zero/span temperature was 25°C; zero/span pressure was 84 kPa; subsequent measurements made at or near the span concentration; temperature is not more than ±6°C from the zero/span temperature; and ambient temperature is within the gas analyzer operating temperature range.
  • 2% (standard deviation of calibration residuals)
Precision RMS (maximum) 0.004 g/m3 (0.006 mmol/mol)

Nominal conditions for precision verification test: 25°C, 86 kPa, 400 μmol/mol CO2, 12°C dewpoint, and 20 Hz bandwidth.
Calibrated Range 0 to 72 mmol/mol (38°C dewpoint)
Zero Drift with Temperature (maximum) ±0.037 g/m3/°C (±0.05 mmol/mol/°C)
Gain Drift with Temperature (maximum) ±0.3% of reading/°C
Cross Sensitivity (maximum) ±0.1 mol H2O/mol CO2

Sonic Anemometer - Accuracy

-NOTE- The accuracy specification for the sonic anemometer is for wind speeds < 30 m s-1 and wind angles between ±170°.
Offset Error
  • < ±8.0 cm s-1 (for ux, uy)
  • < ±4.0 cm s-1 (for uz)
  • ±0.7° while horizontal wind at 1 m s-1 (for wind direction)
Gain Error
  • < ±2% of reading (for wind vector within ±5° of horizontal)
  • < ±3% of reading (for wind vector within ±10° of horizontal)
  • < ±6% of reading (for wind vector within ±20° of horizontal)
Measurement Precision RMS
  • 1 mm s-1 (for ux, uy)
  • 0.5 mm s-1 (for uz)
  • 0.025°C (for sonic temperature)
  • 0.6°C (for wind direction)
Speed of Sound Determined from 3 acoustic paths (corrected for crosswind effects)
Rain Innovative signal processing and transducer wicks considerably improve performance of the anemometer during precipitation events.

Basic Barometer (option -BB)

Total Accuracy
  • ±3.7 kPa at -30°C, falling linearly to ±1.5 kPa at 0°C (-30° to 0°C)
  • ±1.5 kPa (0° to 50°C)
Measurement Rate 10 Hz

Enhanced Barometer (option -EB)

Manufacturer Vaisala PTB110
Total Accuracy ±0.15 kPa (-30° to +50°C)
Measurement Rate 1 Hz

Ambient Temperature

Manufacturer BetaTherm 100K6A1IA
Total Accuracy ±0.15°C (-30° to +50°C)

Compatibilidad

Datalogger Considerations

Compatible Dataloggers

CR200(X) Series CR800/CR850 CR1000 CR3000 CR6 CR9000X CR5000 CR300