Can the CR5000 store data with a rate of 1000 Hz?

On the 5 V range code, with minimal processing, measurement rates up to 2000 Hz on a single channel can be achieved. On the 50 mV range code, the limit is closer to 1000 Hz throughput on a single channel. All 20 differential channels can be measured at a rate of 200 Hz on the 5 V range, and all 40 single-ended channels can be measured at a rate of 125 Hz on a single channel using the 5 V range code.

This was helpful