| Services Available | 
|---|
The SI-111-PW, manufactured by Apogee, is a version of our SI-111 that is fitted with a connector for attachment to a prewired enclosure. This precision infrared radiometer determines the surface temperature of an object without physical contact. The SI-111-PW measures both the subject's surface temperature and the sensor-body temperature. A Campbell Scientific datalogger uses these measurements to calculate the correct temperature of the subject.
Read More
				
							
							The SI-111-PW consists of a thermopile, which measures surface temperature, and a thermistor, which measures sensor body temperature. The two temperature sensors are housed in a rugged aluminum body that contains a germanium window. 
Both the thermopile and the thermistor output a millivolt signal that most of our data loggers can measure. The data logger uses the Stefan-Boltzman equation to correct for the effect of sensor body temperature on the target temperature. The corrected readings yield an absolute accuracy of ±0.2°C from -10° to +65°C.
The SI-111-PW has a 22-degree half-angle field of view (FOV). The FOV is reported as the half-angle of the apex of the cone formed by the target (cone base) and the detector (cone apex). The target is a circle from which 98% of the radiation viewed by the detector is being emitted.
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
| Product | Compatible | Note | 
|---|---|---|
| 21X (retired) | ||
| CR10 (retired) | ||
| CR1000 (retired) | ||
| CR10X (retired) | ||
| CR200X (retired) | ||
| CR206X (retired) | ||
| CR211X (retired) | ||
| CR216X (retired) | ||
| CR23X (retired) | ||
| CR295X (retired) | ||
| CR3000 (retired) | ||
| CR500 (retired) | ||
| CR5000 (retired) | ||
| CR510 (retired) | ||
| CR800 (retired) | ||
| CR850 (retired) | ||
| CR9000 (retired) | ||
| CR9000X (retired) | 
The SI-111-PW is often fastened to a CM200-series crossarm, a tripod or tower mast, or a user-supplied pole using a CM230, CM230XL, or CM220 mount. The CM230 and CM230XL are adjustable inclination mounts that allow the SI-111-PW to be mounted perpendicular to the target surface when the target surface is on an incline. The CM230XL is similar to the CM230, but the CM230XL places the SI-111-PW further from the pole or crossarm. The SI-111-PW may also be attached directly to a user-supplied camera tripod.
The cable on the SI-111-PW is fitted with a connector that attaches to a prewired enclosure.
| Input Power | 2.5 V excitation (for thermistor) | 
| Response Time | < 1 s (to changes in target temperature) | 
| Target Temperature Output Signal | 60 μV per °C difference from sensor body | 
| Body Temperature Output Signal | 0 to 2500 mV | 
| Optics | Germanium lens | 
| Wavelength Range | 8 to 14 μm (corresponds to atmospheric window) | 
| Field of View (FOV) | 22° half angle | 
| Operating Temperature Range | -55° to +80°C | 
| Operating Relative Humidity Range | 0 to 100% RH | 
| Cable Description | 4.5 m (14.76 ft) twisted, shielded 4-conductor wire with Santoprene casing, ending in pigtails | 
| Absolute Accuracy | 
									
										
  | 
							
| Uniformity | 
									
										
  | 
							
| Repeatability | 
									
										
  | 
							
| Diameter | 2.3 cm (0.9 in.) | 
| Length | 6 cm (2.4 in.) | 
| Weight | 190 g (6.7 oz) |